Biometric measurements in highly myopic eyes

Peiyang Shen, Yingfeng Zheng, Xiaohu Ding, Bin Liu, Nathan Congdon, Ian Morgan, Mingguang He*

*Corresponding author for this work

    Research output: Contribution to journalArticlepeer-review

    35 Citations (Scopus)

    Abstract

    Purpose: To assess the repeatability and accuracy of optical biometry (Lenstar LS900 optical low-coherence reflectometry [OLCR] and IOLMaster partial coherence interferometry [PCI]) and applanation ultrasound biometry in highly myopic eyes. Setting: Division of Preventive Ophthalmology, Zhongshan Ophthalmic Center, Guangzhou, China. Design: Comparative evaluation of diagnostic technology. Methods: Biometric measurements were taken in highly myopic subjects with a spherical equivalent (SE) of -6.00 diopters (D) or higher and an axial length (AL) longer than 25.0 mm. Measurements of AL and anterior chamber depth (ACD) obtained by OLCR were compared with those obtained by PCI and applanation A-scan ultrasound. Right eyes were analyzed. Repeatability was evaluated using the coefficient of variation (CoV) and agreement, using Bland-Altman analyses. Results: The mean SE was -11.20 D ± 4.65 (SD). The CoVs for repeated AL measurements using OLCR, PCI, and applanation ultrasound were 0.06%, 0.07%, and 0.20%, respectively. The limits of agreement (LoA) for AL were 0.11 mm between OLCR and PCI, 1.01 mm between OLCR and applanation ultrasound, and 1.03 mm between PCI and ultrasound. The ACD values were 0.29 mm, 0.53 mm, and 0.51 mm, respectively. These repeatability and agreement results were comparable in eyes with extreme myopia (AL ≥27.0 mm) or posterior staphyloma. The mean radius of corneal curvature was similar between OLCR and PCI (7.66 ± 0.24 mm versus 7.64 ± 0.25 mm), with an LoA of 0.12 mm. Conclusion: Optical biometry provided more repeatable and precise measurements of biometric parameters, including AL and ACD, than applanation ultrasound biometry in highly myopic eyes. Financial Disclosure: No author has a financial or proprietary interest in any material or method mentioned.

    Original languageEnglish
    Pages (from-to)180-187
    Number of pages8
    JournalJournal of Cataract and Refractive Surgery
    Volume39
    Issue number2
    DOIs
    Publication statusPublished - Feb 2013

    Fingerprint

    Dive into the research topics of 'Biometric measurements in highly myopic eyes'. Together they form a unique fingerprint.

    Cite this