Preparation of a multi-isotope plutonium AMS standard and preliminary results of a first inter-lab comparison

B. A. Dittmann*, T. J. Dunai, A. Dewald, S. Heinze, C. Feuerstein, E. Strub, L. K. Fifield, M. B. Froehlich, S. G. Tims, A. Wallner, M. Christl

*Corresponding author for this work

    Research output: Contribution to journalArticlepeer-review

    7 Citations (Scopus)

    Abstract

    The motivation of this work is to establish a new multi-isotope plutonium standard for isotopic ratio measurements with accelerator mass spectrometry (AMS), since stocks of existing solutions are declining. To this end, certified reference materials (CRMs) of each of the individual isotopes 239Pu, 240Pu, 242Pu and 244Pu were obtained from JRC IRMM (Joint Research Center Institute for Reference Materials and Measurements). These certified reference materials (IRMM-081a, IRMM-083, IRMM-043 and IRMM-042a) were diluted with nitric acid and mixed to obtain a stock standard solution with an isotopic ratio of approximately 1.0:1.0:1.0:0.1 (239Pu:240Pu:242Pu:244Pu). From this stock solution, samples were prepared for measurement of the plutonium isotopic composition by AMS. These samples have been measured in a round-robin exercise between the AMS facilities at CologneAMS, at the ANU Canberra and ETH Zurich to verify the isotopic ratio and to demonstrate the reproducibility of the measurements. The results show good agreement both between the different AMS measurements and with the gravimetrically determined nominal ratios.

    Original languageEnglish
    Pages (from-to)327-331
    Number of pages5
    JournalNuclear Instruments and Methods in Physics Research, Section B: Beam Interactions with Materials and Atoms
    Volume361
    DOIs
    Publication statusPublished - 15 Oct 2015

    Fingerprint

    Dive into the research topics of 'Preparation of a multi-isotope plutonium AMS standard and preliminary results of a first inter-lab comparison'. Together they form a unique fingerprint.

    Cite this