The Zero Model by using Coincidence Scintillation (ZoMBieS) method of absolute radioactivity measurement

Lindsey J. Bignell, Li Mo, Tristan Steele, S. Reza Hashemi-Nezhad

    Research output: Contribution to journalArticlepeer-review

    3 Citations (Scopus)

    Abstract

    A new method for the absolute radioactivity measurement of radionuclides using a liquid scintillation detector comprising three photomultipliers has been described. The method does not contain any a priori model of the photomultiplier detection efficiency and uses a modified Compton Coincidence technique, therefore it has been named the Zero Model By using Coincidence Scintillation (ZoMBieS) method. Validation measurements of the activity of a (63) Ni source against the Triple to Double Coincidence Ratio method are presented and the results agree within their standard uncertainties.

    Original languageEnglish
    Article number6588983
    Pages (from-to)4007-4014
    Number of pages8
    JournalIEEE Transactions on Nuclear Science
    Volume60
    Issue number5
    DOIs
    Publication statusPublished - 2013

    Fingerprint

    Dive into the research topics of 'The Zero Model by using Coincidence Scintillation (ZoMBieS) method of absolute radioactivity measurement'. Together they form a unique fingerprint.

    Cite this